Single Axis Fluxgate Magnetometers

Mag-01 and Mag-01H
These portable high performance instruments provide precision
measurements of the direction and intensity of static and slowly varying
magnetic fields from 0.1nT to 2mT (1mT=10G). Axial and transverse probes
are available together with unpackaged probes for cryogenic applications.
Both instruments are mains or battery powered and have a 4½ digit
display and analogue output. The internal rechargeable battery provides
16 hours continuous use and can be recharged using the mains adaptor
provided.
The Mag-01 provides a maximum resolution of 1nT whilst the Mag-01H
has an additional sensitivity control providing a resolution of 0.1nT
and an offset control for up to ±90µT in 10µT steps.
These instruments feature superb linearity and accuracy and very low
drift with time and temperature. Calibration is traceable to international
standards and a re-calibration service is available.
Applications include the measurement of induced and remanent magnetism
and use in site surveys prior to the installation of MRI equipment.
Mag Probes
The following probes are available:
| Type |
Low
field probes (0 to 0.2mT) |
High
field probes (0 to 2mT) |
| Axial |
Mag B Probe |
Mag D Probe |
| Transverse |
Mag C Probe |
Mag E Probe |
| Cryogenic |
Mag F Probe |
Mag G Probe |
The measurement range/resolution (LCD display) are:
| Magnetometer |
Low field probes |
High field probes |
| Range(µT) |
Resolution(nT) |
Range(µT) |
Resolution(nT) |
| Mag-01 |
0-20 |
1 |
0-200 |
10 |
| 20-200 |
10 |
200-2000 |
100 |
| Mag-01H |
0-2 |
0.1 |
0-20 |
2 |
| 2-100 |
1 |
20-1000 |
10 |
| 100-290 |
10 |
1000-2000 |
100 |
The sensing axis of the B and D probes is in line with the axis of
the cylindrical enclosure. The sensing axis of the C and E probes is
along the axis of the short cylinder at the end of the probe. The magnetic
detection axis and centre of sensitive volume are marked on each probe.
The cryogenic probes are suitable for operation at temperatures down
to liquid helium and have 1m of 4 x 0.2mm enamelled copper wire connected
to the element for use in a cryostat or vacuum chamber. This length
can be extended to a maximum of 5 metres if required.
A magnetometer probe switch unit is available for the sequential selection
of two or three probes, type B-G, for operation with a Mag-01 or Mag-01H
instrument. The Mag probes can be operated at a distance up to 25 metres
from the switch unit.
Mag-01
and Mag-01H Electronics Unit Specification |
Performance
|
Measuring range |
0 to 0.2mT or 2mT depending on probe
|
Bandwidth x1 sensitivity (low field probes)
|
dc to -3dB at 10Hz, 20µT P-P, -12dB per octave roll
off |
Calibration accuracy |
0.1% |
Maximum resolution |
1nT (Mag-01), 0.1nT (Mag-01H) |
Zero field offset |
±5nT |
Offset drift |
0.01nT/°C |
Calibration temperature coefficient |
<10ppm/°C |
Liquid crystal display |
4½ digit autoranging |
For high field probes multiply reading
on display by 10 to obtain a true value |
Front panel controls/sockets
|
On/Off switch |
switches on the internal battery supply
|
Probe input |
6 pole waterproof Fischer connector |
Charge indicator |
LED illuminated when external supply is connected
|
Offset control
(Mag-01H only) |
allows a preselected field strength between ±90µT
in ±10µT steps to be subtracted from the field experienced by
the probe |
Sensitivity control
(Mag-01H only) |
increases the sensitivity of the magnetometer by
a factor of 10 |
Rear panel sockets
|
Battery charger inlet |
2.1mm socket dc 0.5A max. Polarity protected 6-18V
continuous or intermittent use |
Analog output
- low field probes
- high field probes |
2 x 4mm rear panel insulated sockets
5V full scale 100µT/V (10µT/V with x10 sensitivity on Mag-01H)
2V full scale 1mT/V (0.1mT/V with x10 sensitivity on Mag-01H)
|
Output impedance |
1 kohm |
Mechanical
|
Dimensions |
155 x 170 x 68mm |
Enclosure |
high impact ABS |
Weight |
1.5kg |
Environmental
|
Operating temperature range |
-10°C to +50°C |
Relative humidity |
80% maximum non-condensing |
Accessories
|
Instrument stand, carrying bag
|
Mains adaptor/battery charger - 110V
or 220V (as requested) |
Option - service manual |
Mag Probes Technical Specification |
Calibration accuracy |
1% |
Linearity |
0.01% |
Connection cable |
4 core screened 5 metre with integral 6 pole Fischer
connector. Alternative lengths of cable (maximum length 25 metres)
are available on request |
Environmental sealing |
submersible in water to 10 metres |
Shock resistance |
drop tested from 0.5 metre onto PVC floor
3 times |
Enclosure material |
gold plated brass |
Probe alignment to case |
better than 0.2 degrees |
Operating temperature range
- axial and transverse probes
- cryogenic probes |
-30°C to +75°C
Liquid helium to +75°C |
Low field Mag
Probes type B, C and F
|
Temperature coefficient of calibration
|
±10ppm/°C over full range of operating temperatures
|
Offset error |
<1nT over full range of operating temperatures
|
Sensitive volume |
0.02cm³ |
Excitation power |
26mW |
High field Mag
Probes type D, E and G
|
Temperature coefficient of calibration
|
±30ppm/°C over full range of operating temperatures
|
Offset error |
<20nT over full range of operating temperatures
|
Sensitivity volume |
0.0035cm³ |
Excitation power |
16mW |
Probe
Construction |
![[Drawing]](safMagnetometers/bart014t.gif) |
![[Drawing]](safMagnetometers/bart015t.gif) |
![[Drawing]](safMagnetometers/bart016t.gif) |
| Axial
|
Transverse
|
Low
Temperature
|